Dimens 420 Dissolved Oxygen Sensor
Discover the Dimens 420 Dissolved Oxygen Sensor: Perfect for Water Applications
Applications: Surface water, sewage treatment, and water plant farming.
Features: Prepare for accurate measurements with easy assembly and maintenance. Ensure tight connections for reliable results.
Specifications:
- Type: Polarographic Dissolved Oxygen Electrode
- Material: Stainless steel, PEEK, silicone, NBR
- Measuring Range: 0-20 ppm
- Operating Temperature: 0-60°C
- Pressure: 0-4 bar
- Flow Rate: Minimum 0.03 m/s
- Cable Length: 5m
- Protection Class: IP68
Enhance your water quality monitoring with Dimens
Proper Setup Steps:
- Unscrew the dissolved oxygen membrane cap from the electrode column.
- Rinse the internal cathode/anode with distilled water, then dry it.
- Add the provided solution to the cap and attach it to the column; don't worry if some spills during assembly.
- Secure the cap tightly, ensuring the cathode aligns with the membrane (no folds or tears).
- Rinse the electrode, place it in the sheath, and prepare for use. Note: After the first use or membrane replacement, polarization is necessary and takes about 15 minutes."